Acteon MINILED SUPERCHARGED - Performing a Cure Cycle; Switching off the Medical Device; Disconnecting the Medical Device

Acteon MINILED SUPERCHARGED
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mode Exposure time
Fast
3seconds
4seconds
5seconds
10seconds
Pulsed
5 1-second bursts, with 250-millisecond intervals between each burst
10 1-second bursts, with 250-millisecond intervals between each burst
Progressive
6 seconds progressive and 3 seconds at full power
10 seconds progressive and 5 seconds at full power
Available exposure times
6.4.3 Performing a cure cycle
1. Position the end of the optical guide 2mm from the surface of the composite to be cured. Do not touch it.
2.
Press the button to start the cure cycle.
A beep sound confirms start of the cycle. During the cycle, the remaining cycle time counts down on the
MINILED LCD screen.
A second beep sound confirms cycle completion.
Depending on the cure composite used, repeat the cure cycle as required.
6.5 Switching off the medical device
1.
Press the button to stop a cure cycle.
2. Replace the MINILED on the charging base after each use for optimised battery efficiency.
After 5 minutes of inactivity, the MINILED switches to standby and the on/off light indicator goes off.
6.6 Disconnecting the medical device
Before a long absence or when not in use, the medical device must be cleaned, its battery must be removed and the
charging base must be disconnected from the mains power.
Page 15/38 - User Manual|MINILED SUPERCHARGED|J02261EN|V2|(09)|08/2018|NG35EN010B