Possible causes Solutions
The tip is incorrectly tightened
Fasten the tip using the wrench
Replace your torque wrench once a year
Faulty connector contact Clean the cord contacts
Cut handpiece cord
Return the handpiece and cord to the Acteon Customer
Service team to have them replaced
Adjust the power Please read the Setting the power chapter
10.5 Water leakage
Symptoms: There is a water leak along the irrigation line or at the handpiece and cord assembly.
Possible causes Solutions
Dysfunctional irrigation line Replace the irrigation line with a new one.
10.6 Disturbance of settings defined by the practitioner
Symptoms: the medical device suddenly switches from one mode to another, the irrigation flow increases/decreases or
the purge function suddenly starts up.
Possible causes Solutions
Liquid present on the medical device screen
Switch off the medical device and wipe with a clean and
dry cloth
Page 28/39 - User Manual | Piezotome Cube | J50101 | V4 | (17) | 03/2018 | NO37EN010D