Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
91 B
"B_AXIS_ERR_RUN_PTPT
Unexpected FSM condition during
B-axis position adjustment"
A
Reset error from workstation; if error remains restart trium and
workstation.
92 C
"R_AXIS_ERR_RUN_LIMI
R-axis position out of range"
A
Reset error from workstation; if error remains restart trium and
workstation.
93 C
"HI_VOLT_ENA_IN_IDLE
XRay button activated when not
allowed"
U/T
"Reset error from workstation; if error remains check that XRay
exposure buttoni s not pressed; If error remains technical
service is required:
check if XRay exposure button is shorted."
94 C
"INVERTER_LINK_ERROR
CAN communication error between
Mainboard and Inverter"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check CAN cables between Maiboard and Inverter board;
check/replace Mainboard;
check/replace Inverter board."
95 B
"FW_UPDATED_CANT_EXE
Cannot exec rmware after update
(board need hardware reset)"
T
Reset error from workstation; if error remains restart trium and
workstation.
96 C
"EEPROM_DEVICE_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 0 -
EEPROM"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
97 C
"EEPROM_DMA_RD_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 1 -
EEPROM READ DMA FAILURE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
98 C
"EEPROM_DMA_WR_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 2 -
EEPROM WRITE DMA FAILURE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
99 C
"EEPROM_I2CBUS_ERROR
Inverter board error:
GENERAL STATE ERROR, bit 3 -
EEPROM I2CBUS"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
100 C
"ANOD_CURR_NOT_CALIB
Inverter board error:
GENERAL STATE ERROR, bit 4 -
ANODIC CURRENT NOT CALIBRATED"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."