41
When using AC adapter, the calibrator cannot auto power off.
This function will be invalid when performing Pressure control, Auto step or Task.
The Auto power off will be invalid when the screen saver is set to Never.
3.4.5 Charge Mode
The device provides fast charge mode and high precision charge mode. The quick charge mode can improve
charging speed significantly, but it has influence to the accuracy for pressure measurement. The high precision mode can
ensure the accuracy for measurement but will slow down the charge speed.
The default mode is fast charge mode when device is turned on. Select high precision mode if you need to perform
high accuracy pressure measurement during charging through adapter, but this will not be saved and will be reset to the
quick charge mode after the device is restarted.
3.5 Product Information
Product information is read only, it includes basic information, version and running information:
Basic information: model, serial number and scale information.
Version: Host, system version, pressure module version and Wi-Fi version. The firmware version is usually referred
to Host version. When contact with the customer service team, please provide information of Host version if needed.