Example: C|20120510|112632|Advanced Instruments 3320|
03090845A|20060510|80000|1|1|1 <CR> <LF>
Message Type: ERROR
The ERROR message is issued anytime the firmware issues a Send-
Error() command.
The message is initiated using the uppercase letter ‘E’, contains a
delimiter between each segment, and is terminated with CR/LF. The
contents of the message include the message type, date, time, com-
pany name, model number, serial number, sample ID if any, error
code, and error message text.
Example: E|20120510|112632|Advanced Instruments |3320|
03090845A|0123456789ABCDEFGHIJ|1000|Sample Pre Freeze
<CR> <LF>
Data Collection
Basic Tools
AILink
The Advanced Instruments utility, AILink may be used to collect and
save the serial port messages via the AITerminal tab. This will cre-
ate a text file that may be imported into third party software, such as
Microsoft
®
Excel. Refer to Excel documentation on how to import a
delimited text file. Once properly imported, you may use the head-
ing information supplied above for each message type to read or sort
the message strings.
HyperTerminal
Some versions of Microsoft Windows
®
contain a utility for serial
port communications that can be used to capture serial port output in
much the same manner as described for AILink. Consult Microsoft
®
documentation and the instrument users guide for information on
setting up a HyperTerminal session.