Chapter 4 Diagnostics
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
22
4.4 Testing the sample probe
1. Press DIAGNOSTICS > SAMPLE PROBE to open the
Sample Probe diagnostics screen.
You are prompted to test a sample of deionized
water; the test results display in Ohms on this screen.
NOTE: The result should be close in value to the
Sample Probe Resistance value set in
Changing the probe resistance.
Test progress shows in the indicator while the test is
in progress.
2. Press Exit to return to the Home screen.
4.5 Testing the barcode scanner
The following barcodes are recognized by the instrument:
Code 39 Interleaved
Code 128 PDF417
ECC MSI
UPC QR
GS1 IATA
Codabar MaxiCode
Code 93 Aztec
Code 11 EAN
1. Select DIAGNOSTICS > BARCODE to open the
Barcode diagnostics screen.
2. To test if the barcode scanner is working properly,
place any barcode of a supported type a few inches
in front of the barcode scanner.
If successful, you hear a beep and the barcode
information displays on screen.
3. Press Exit to return to the Diagnostics menu.
4.6 Viewing and exporting events
An event is defined as any action that occurs with the
instrument, including but not limited to:
System errors
Settings changes
User added or deleted
Power up
Controls added or deleted
Calibration results
Consumables box (test kit) not placed correctly
Cleaning/preventative maintenance performed
Test results reviewed/approved
In a busy lab, this list might contain several hundred
events. Users can scroll or navigate through the Events
log using the icons at the top of the screen, create a
filter to narrow to events of interest, and/or export the
events in the configured file format and location.
See Configuring data export to configure the location
where data is exported (USB drive or network folder)
and file format(s) of the exported data (PDF and/or
CSV).
1. Press DIAGNOSTICS > EVENTS to open the Events
log.