Chapter 5 Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
30
Configuring OPC-UA communication
NOTE: For more information about OPC-UA, see OPC-UA
Features and Configuration.
1. Press NETWORKING and log in.
2. Press OPC-UA.
3. Press Enable.
4. Enter the server port, device name, and minimum
sample interval (in msec).
5. Press SAV E.
6. Power o the instrument and wait 15seconds for the
OPC-UA change to take aect.
7. Power on the instrument.
Enabling the web server
Use the embedded web server to view and print
information—such as results, status, and data logs—from
a network computer or mobile device.
1. Connect the instrument to the network using an
Ethernet cable.
2. Press NETWORKING and log in.
3. Press the WEB SERVER tab.
4. Select Web Server Enabled.
5. Press to change the number of records (5-1000)
to display and press SAV E.
6. Press and select LAN to display the web server IP
address.
7. Enter the IP address into the address bar of any web
browser to display, save, and print the data.