Chapter 4 Diagnostics
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
25
Logging routine instrument cleaning
1. Press DIAGNOSTICS > MAINTENANCE and log in.
2. Press the CLEANING tab.
3. Select Cleaned Sample Probe and/or Cleaned
Chamber if those tasks were completed
4. Press Save.
5. Press OK on the Cleaning completed dialog box.
6. Press Exit to return to the Diagnostics menu.
Logging preventative maintenance
Administrator-level users and Service account users can
record when preventative maintenance is performed
and the service engineer performing the maintenance.
The name of the service engineer is required for all
preventive maintenance events.
Refer to Maintenance and service for instrument
maintenance recommendations and instructions.
To record a maintenance event, complete the following
steps.
1. Press DIAGNOSTICS > MAINTENANCE and log in.
2. Press the PREVENTATIVE MAINTENANCE tab.
3. Select the Advanced Instruments Preventative
Maintenance checkbox to record the maintenance event.
4. Press to enter the name of the Advanced
Instruments trained service engineer (required)
who performed the maintenance.
5. Press EDIT to type a Comment (required - up to
1024 characters) to enter a full explanation of the
maintenance performed.
6. Press Save.
The instrument saves the maintenance information
to the events database and updates the preventative
maintenance date. The instrument displays prompts
starting 11months after this date as a reminder that
the next recommended maintenance is due.
NOTE: To set the installation date (and initial
preventative maintenance date) for a new
instrument, refer to Setting the installation date.
7. Press Exit to return to the Diagnostics menu.