Chapter 5 Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
28
5. Enter a date and time format:
• Select Use Regional Format Settings to
automatically set the date and time formats based
on the language settings (see Changing the
language).
• Otherwise, enter the Date , Year, and Time formats,
and the Delimiter as needed.
Entering the date and time manually
To enter the date and time manually (without
synchronizing to a server):
1. Press DATE & TIME and log in.
2. Clear the Use Time Server checkbox if it is selected.
3. Enter the Date.
4. Enter the Time.
5. Select Use Regional Format Settings checkbox to
automatically set the date and time formats based on
the language settings (see Configuring testing access)
and time format.
Otherwise, clear the Use Regional Format Settings
checkbox and enter the Date , Year, and Time
formats, and the Delimiter as needed.
6. Press SAV E .
5.3 Changing the language
The language is set to English at the factory. You can
change it to one of nine other languages.
1. Press LANGUAGE and log in.
2. Select a language and press SAVE.
3. Press Exit to return to the Settings menu.
NOTE: If the system detects unsaved changes, it
prompts you to save the changes before
exiting.
4. Power o the OsmoTECH XT and wait 15seconds for
the language change to take aect.
5. Power on the instrument.
5.4 Changing the probe resistance
Only change the probe resistance setting after replacing
the sample probe or block probe.
After changing a probe resistance setting, you must
perform a thermal module setup and then recalibrate the
instrument.
1. Press PROBE RESISTANCE and log in to open the
PROBE RESISTANCE screen.