OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
50
Appendix B: Product specifications
Sample type
Aqueous based solution
Units of measure
mOsm/kg H
2
O
Sample volume
20 ± 1 μL
Sample capacity
Single sample
Test time (single sample)
150 seconds (typical) for low range
1
190 seconds (typical) for high range
1
Power requirement
100 to 240 VAC (50/60 Hz) / 0.5 A
Power consumption: 60 W max.
Fuses (2): 250 V time delay (Type T) 1-Amp
Memory backup
Integral lithium cell; 10-year life (unpowered); not user-replaceable
Classifications
Protection class: I
Over-voltage category: II
Pollution degree: 2
Moisture protection: IPX0 (ordinary)
Communications
USB 2.0 Type A ports (2)
USB 2.0 Type B ports (1)
Ethernet 10/100, RJ45 connector port (1)
Operating conditions
Temperature: 18 °C to 35 °C (64 °F to 95 °F)
Humidity: 5% to 80% relative humidity (non-condensing)
Storage temperature
-20 °C to 45 °C (-4 °F to 113 °F)
Dimensions (D x W x H)
Not including test kit: 15" x 14" x 10" (38 cm x 36 cm x 25 cm)
Including test kit: 15" x 14" x 11.5" (38 cm x 36 cm x 38 cm)
Net weight
5.9 kg (12.9 lbs.)
Shipping weight
11.4 kg (25 lbs.)
Warranty
One-year limited warranty on workmanship and parts
Certifications
1 Performance at Reference Conditions: 20 °C to 25 °C (68 °F to 77 °F); 40 to 60% relative humidity.