Chapter 5 Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
41
Resetting instrument system settings
To change the serial number, perform a factory reset,
and/or perform a system reset, complete the following
steps:
1. Select SYSTEM UPDATE and log in to the Service
account.
2. Press the System Settings tab.
3. To change the serial number:
a. Press the Serial Number tab.
b. Press , enter the serial number, and press Enter.
c. Press SAVE.
4. To reset the instrument to factory settings:
a. Press the Factory Reset tab.
b. Press RESET.
5. To perform a system reset (factory reset and also
clears calibration, resistance, alignment, and serial
number):
a. Press the System Reset tab.
b. Press RESET.
Configuring V&V settings
To enable system screen captures, short system
timeouts, and disable calibration requirements,
complete the following steps:
1. Select SYSTEM UPDATE and log in to the Service
account.
2. Press the V&V tab.
3. To enable short timeouts, select the Use short
timeouts checkbox.
4. To bypass Calibration required errors and allow
service engineers to perform tests, select the Ignore
Calibration Required checkbox.
This checkbox is reset (cleared) when the
instrument is powered down and then powered up.
5. Press SAV E.