EasyManua.ls Logo

Advanced Instruments OsmoTECH XT - Setting up Networking Communications; Configuring a LAN Connection

Advanced Instruments OsmoTECH XT
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Configuring settings
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
29
2. On the Probe Resistance Information tab, press
next to the resistance field you want to change for
each block and sample probe.
R0 - Set the sample and/or block probe resistance
at 0 °C to a value between to 5510 and 6090.
R-10 - Set the sample and/or block probe
resistance at -10C to a value between to 8376 and
9712.
3. Press SAVE and then follow the onscreen
instructions.
The changed resistance settings are saved and
logged to the Events database.
4. Power o the instrument and wait 15seconds.
5. Power on the instrument.
6. From the Home screen, press > SETTINGS >
PROBE RESISTANCE.
7. Press the Thermal Module Setup tab and follow the
instructions on the screen.
Once the testing is complete, the slope and intercept
values are saved to the system firmware.
8. Recalibrate the instrument from the Home screen
(see Calibration).
NOTE: Open calibration standards before use. Do not
open them in advance.
5.5 Setting up networking communications
A network administrator can help you set up or
troubleshoot the instrument to connect to a network in
order to send data, back up files, or use a network time
server.
Configuring a LAN connection
The instrument can connect to a computer or other
devices over a local area network (LAN).
1. Plug an Ethernet cable into the Ethernet connector
(see Setting up networking communications).
2. Press NETWORKING and log in.
3. Press the LAN tab.
4. Select one of the following network protocols:
Static IP (recommended for best results): Assigning
an unchanging static IP address to the instrument
makes it easier for other devices to contact the
instrument over the network.
a. Press and enter the IP Address.
b. Press and enter the Subnet Mask.
c. Press and enter the Default Gateway.
d. Press and enter the DNS Server.
DHCP (Dynamic Host Configuration Protocol): A
DHCP server dynamically assigns an IP address and
other network parameters to the instrument.
The IP Address, Subnet Mask, and Default Gateway
are entered automatically, and can change each time
the instrument connects to the network.
5. Enter the host name (up to 32 characters)—
especially if more than one OsmoTECH XT is
connected to the network.
6. Press Save.

Table of Contents

Other manuals for Advanced Instruments OsmoTECH XT

Related product manuals