Ground Resistance Tester Model 6471
CHAPTER 3
SPECIFICATIONS
3.1 Reference Conditions
Influencing Parameters Reference Values
Temperature 68°F ± 5.5°F (20°C ± 3°C)
Relative Humidity 45 to 55% RH
Power Supply 9 to 11.2V
Frequency Range of Input Signal 0 to 440Hz
Capacitance Parallel to Input Resistance 0 uF
Electric Field <1 V/m
Magnetic Field < 40 A/m
3.2 Electrical
3.2.1 Frequency Measurements
Measuring method: Digital with a sampling frequency of 4028Hz, low pass, FFT.
The frequency of the strongest spectral component is displayed.
Measuring rate: Display updates approx. 3 times per s.
Measurement Range 5 to 450Hz
Resolution 1Hz
Operating error ± 2Hz
Min. input voltage 10mV
Min. current through a clamp 0.5mA
3.2.2 Voltage Measurements
Overvoltages up to 75Vrms are displayed as ">65V". Permanent overvoltages
between 70 and 75V at terminals H (Z) and E (X) can cause overheating of the
overvoltage protection varistor.
Voltages of more than 75Vrms lead to error message 31 (excessive external
voltage) or 32 (voltage measurement overrange).
If terminals H (Z) and E (X) are put into contact with the line voltage, the protection
fuse will blow.