Ground Resistance Tester Model 6471
If you need to change the Object number, press the ▲ button until the desired
number is displayed. Then press the ► button; the Test number now blinks.
Press the ▲ button until the desired test is displayed.
Press the MEM button to display the dEL screen. By default, the blinking word
NO appears on the screen, indicating the test is not to be deleted. To delete
the test, press the ▲ button to change the blinking NO to YES, then press and
hold down (>2s) the MEM button. To exit without erasing, perform a short press
on the MEM button.
Note that if the test stored in this Object/Test number is a sweep test (which con-
sists of multiple measurements taken at different frequencies) this procedure only
deletes a single measurement. The remaining sweep measurements will still be
stored in this Object/Test. You must delete all individual measurements to com-
pletely remove a sweep test.
When a test is deleted, its Object/Test number is removed from memory. When you
subsequently view stored tests, this number will be skipped. This Object/Test num-
ber combination remains unavailable for storing tests until you completely erase
memory, as instructed in Section 7.3.1.