Appendix G
Subject to Export Control, see Cover Page for details.
G-5
Device Specific Errors
Device Specific Errors generate an immediate error message to warn users when the Test Set
experiences an internal fault that results in operational failure or causes the Test Set to operate
outside of product specifications. Persistent Device Specific Error messages are displayed until
the cause of the error is resolved.
Error Number 301
“User calibration required"
Ambient Temperature has exceeded limit for current User Calibration setting. Run User
Calibration or change User Calibration Threshold on User Calibration Utils Tile.
Error Number -310
"System error: Measurement processor halted"
Indicates internal hardware malfunction. Reboot Test Set. Contact Aeroflex Customer Service if
error persists.
“System error: Protocol Processor halted”
Indicates internal hardware malfunction. Reboot Test Set. Contact Aeroflex Customer Service if
error persists.
"System error: Transmit processor halted"
Indicates internal hardware malfunction. Reboot Test Set. Contact Aeroflex Customer Service if
error persists.
"System error: Receive processor halted"
Indicates internal hardware malfunction. Reboot Test Set. Contact Aeroflex Customer Service if
error persists.
"System error: Reference is UNLOCKED"
Test Set is not locked to reference. If External is selected, 10 MHz source must be present on
External Reference I/O Connector.
If Internal is selected and error message is received, contact Aeroflex Customer Service.
"System error: Receiver synthesizers are UNLOCKED"
Test Set is not locked to reference. If External is selected, 10 MHz source must be present on
External Reference I/O Connector.
If Internal is selected and error message is received, contact Aeroflex Customer Service.
“System error: Generator synthesizers are UNLOCKED”
Test Set is not locked to reference. If External is selected, 10 MHz source must be present on
External Reference I/O Connector.
If Internal is selected and error message is received, contact Aeroflex Customer Service.
Error Number -313
"User Calibration memory lost"
Test Set lost internal calibration values. Re-Run user calibration. If problem persists contact
Aeroflex Customer Service.
Error Number -321
"Out of memory"
Indicates internal hardware malfunction. Reboot Test Set. Contact Aeroflex Customer Service if
error persists.
Error Number -340
"Calibration failed"
Received when User Calibration fails. Restore Factory defaults and re-run User Calibration. If
fault continues, contact Aeroflex Customer Service.