Agilent 1260 Infinity II Preparative Autosampler User Manual 113
Troubleshooting and Diagnostics
7
Maintenance Functions
Change Metering Seal or Piston
The change-metering seal/piston function draws the piston away from the
home position, relieving the tension on the spring. In this position, the
analytical head assembly can be removed and reinstalled easily after
maintenance.
User Interface
The commands for the control software are:
Start
Draws the piston away from the home position, relieving the tension on the
spring.
Next
Repositions the piston at the home position.
Back
Completes the procedure.
Using the Change Seal Function
1 Select Start to move the piston to the maintenance position.
2 Exchange the metering seal (see “Exchange the Metering Seal and
Piston” on page 173).
3 Select Next to move the piston back to the home position.
4 Select Back to finish the procedure.
NOTE
Read carefully the description in the LabAdvisor user interface.
NOTE
Read carefully the description in the LabAdvisor user interface.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals