Agilent 1260 Infinity II Preparative Autosampler User Manual 173
Maintenance
9
Exchange the Metering Seal and Piston
Exchange the Metering Seal and Piston
When Poor injection-volume reproducibility
Leaking metering device
Tools required p/n Description
8710-0510 Wrench open 1/4 — 5/16 inch
8710-2392 4 mm Hex key
8710-2411 Hex key 3 mm12 cm long
01018-23702 Insert tool
Cleaning tissue and appropriate solvent like isopropanol or methanol
Parts required p/n Description
0905-1294 Metering seal, 900 µL
G4267-60462 Piston Assembly, 900 µL
Preparations Finish any pending acquisition job and in order to avoid leaks, stop the pump running.
1
In the Tools section of the Agilent Lab Advisor software
select Service & Diagnostics > Maintenance Positions
> Change Metering Device, click Start and wait until
the metering device is in maintenance position.
2
Open the front door.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals