EasyManua.ls Logo

Agilent Technologies 1260 Infinity II - Swage Fittings

Agilent Technologies 1260 Infinity II
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity II Preparative Autosampler User Manual 83
Preparing the Module
5
Swage Fittings
Swage Fittings
1
Select a nut that is long enough for the fitting you'll be
using.
2
Slide the nut over the end of the tubing.
3
Carefully slide the ferrule components on after the nut
and then finger-tighten the assembly while ensuring that
the tubing is completely seated in the bottom of the end
fitting.
4
Use a column or injection valve to gently tighten the
fitting which forces the ferrule to seat onto the tubing.
NOTE
Don't overtighten. Overtightening will shorten the
lifetime of the fitting.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Related product manuals