EasyManuals Logo

Agilent Technologies 1260 Infinity II User Manual

Agilent Technologies 1260 Infinity II
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
130 Agilent 1260 Infinity II Preparative Autosampler User Manual
8
Error Information
General Error Messages
General Error Messages
Timeout
Error ID: 0062
The timeout threshold was exceeded.
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable cause Suggested actions
1
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
2
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Probable cause Suggested actions
1
Leak detected in another module with a CAN
connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down
condition.
4
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals