Agilent 1260 Infinity II Preparative Autosampler User Manual 179
Maintenance
9
Replace the Peristaltic Pump Cartridge
Replace the Peristaltic Pump Cartridge
When Tubing blocked or broken
Tools required Description
Parts required p/n Description
5065-4445 Peristaltic pump with Pharmed tubing
Preparations Remove the inlet filter of the solvent bottle which guides the solvent to the peristaltic pump to
avoid syphoning effects.
WARNING
When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can hold health risks.
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
NOTE
The peristaltic pump cartridge is a replaceable unit. The tubing inside the pump is not
replaceable.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals