Agilent 1260 Infinity II Preparative Autosampler User Manual 179
Maintenance
9
Replace the Peristaltic Pump Cartridge
Replace the Peristaltic Pump Cartridge
When Tubing blocked or broken
Tools required Description
Parts required p/n Description
5065-4445 Peristaltic pump with Pharmed tubing
Preparations Remove the inlet filter of the solvent bottle which guides the solvent to the peristaltic pump to
avoid syphoning effects.
When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can hold health risks.
➔ Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
The peristaltic pump cartridge is a replaceable unit. The tubing inside the pump is not
replaceable.