EasyManua.ls Logo

Agilent Technologies 1260 Infinity II - Page 160

Agilent Technologies 1260 Infinity II
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
160 Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Needle Seat Assembly
3
Unscrew the safety cover screw.
WARNING
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
4
Remove the safety cover.
WARNING
Sharp needle
Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
5
Turn or flap the wash port to the left.
6
In Lab Advisor use Next to move the needle in the down
position.
The needle will move slowly downwards and will stop
approximately 2 mm above the seat.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Related product manuals