166 Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Sample Loop Assembly
7
Reinstall the new sample loop.
NOTE
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
8
In Lab Advisor use Next to lift the needle slowly into the
up position.
9
Turn or flap the wash port to the right.
10
Install the safety cover (1) and fix the screw (2).
11
Click Next until the needle assembly is moved to the
needle seat.
12
Select Back to finish the procedure.
13
Close the doors.
14
Perform a pressure test.