184 Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Wash Port Assembly
9
Loosen the fixing screw.
10
Lift out the needle.
11
Disconnect the wash tubing from the peristaltic pump.
12
Release the spring loaded pin.
13
Lift the flapper out of the bracket. Be careful not to
squeeze the wash tubing.
14
Reconnect the wash tubing fitting to the new wash port
assembly.