186 Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Wash Port Assembly
19
In Lab Advisor use Next to lift the needle slowly into the
up position.
20
Turn or flap the wash port to the right.
21
Install the safety cover (1) and fix the screw (2).
22
Select Next to move the needle into the needle seat.
23
Select Back to finish the procedure.
24
Close the doors.
25
Purge the wash port. Check if solvent will be delivered
from the peristaltic pump.