62 Agilent 1260 Infinity II Preparative Autosampler User Manual
4
Installation of the Optional Sample Cooler
Install the Sample Cooler
5
Slide in the sample cooler the halfway.
Module is partially energized when switched off, as long 
as the power cord is plugged in.
Repair work at the module can lead to personal injuries, 
e.g. shock hazard, when the cover is opened and the 
module is connected to power. 
➔
Make sure that it is always possible to access the 
power plug.
➔
Do not use the sample cooler if it is not operating 
correctly or has been damaged. Disconnect it from the 
power supply and call your local service center.
➔
Remove the power cable from the module before 
opening the cover.
➔
Do not connect the power cable to the module while 
the covers are removed.
➔
If the sample cooler is disconnected from the power 
supply, you should wait for at least five minutes before 
switching on the compressor.
Damaged electronics
➔
To avoid damages of the electronics of the module 
make sure the power cords are unplugged before 
disconnecting or reconnecting the sampler to the 
sample cooler cables.
6
Connect power cable and signal/data cable.