62 Agilent 1260 Infinity II Preparative Autosampler User Manual
4
Installation of the Optional Sample Cooler
Install the Sample Cooler
5
Slide in the sample cooler the halfway.
WARNING
Module is partially energized when switched off, as long
as the power cord is plugged in.
Repair work at the module can lead to personal injuries,
e.g. shock hazard, when the cover is opened and the
module is connected to power.
Make sure that it is always possible to access the
power plug.
Do not use the sample cooler if it is not operating
correctly or has been damaged. Disconnect it from the
power supply and call your local service center.
Remove the power cable from the module before
opening the cover.
Do not connect the power cable to the module while
the covers are removed.
If the sample cooler is disconnected from the power
supply, you should wait for at least five minutes before
switching on the compressor.
CAUTION
Damaged electronics
To avoid damages of the electronics of the module
make sure the power cords are unplugged before
disconnecting or reconnecting the sampler to the
sample cooler cables.
6
Connect power cable and signal/data cable.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals