64 Agilent 1260 Infinity II Preparative Autosampler User Manual
4
Installation of the Optional Sample Cooler
Install the Sample Cooler
11 12
NOTE
Check leak waste handling for further info.
CAUTION
Damage to the sample cooler
Wait at least 30 min before switching on the
compressor of the sample cooler.
This allows the refrigerant and system lubrication to
reach equilibrium.
13
Connect the power cable to the power connector at the
rear of the module.