EasyManuals Logo

Agilent Technologies 2100 User Manual

Agilent Technologies 2100
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
Contents 194 Index
4 Carefully close the lid.
CAUTION
Do not force the lid closed. This can damage the pressure cartridge. If the lid does not
close without force, check that chip is inserted correctly and that the chip selector is
at the correct position for this chip type.
When the software recognizes an inserted chip, the chip is shown on the Instrument
tab. If you have closed the lid, and the software has not recognized the chip, verify that
the cartridge and chip are inserted properly and the chip selector is in the correct
position. Close the lid.
The adapter with the gasket in the cartridge fits onto the priming well of the chip. A
small gap between the lid and the instrument mainframe is normal and no cause for
malfunction. The icon on the Instrument tab changes to a cell chip icon:
If the chip is not detected, open and close the lid again.

Other manuals for Agilent Technologies 2100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 2100 and is the answer not in the manual?

Agilent Technologies 2100 Specifications

General IconGeneral
BrandAgilent Technologies
Model2100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals