EasyManua.ls Logo

Agilent Technologies 2100 - Page 22

Agilent Technologies 2100
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Contents 22 Index
When the chip is detected, the image on the Instrument tab changes to a chip.
If the chip is not detected, open and close the lid again.
NOTE
If the AutoRun option is active, the chip run starts automatically once a chip has been
inserted and the lid has been closed.
6 On the Instrument tab, click the Start button.
The chip run starts. The Raw Signals sub-tab shows an electropherogram of the currently
measured sample. The name of the sample is displayed above the graph. The graph is a
“live” plot of the migration time against fluorescence units (raw data, including
background fluorescence, for example).

Other manuals for Agilent Technologies 2100

Related product manuals