System Verification
System verification proves that the bioanalyzer system is suitable for its intended use,
that is, that it will function according to its operational specifications in the selected
environment.
System verification should be performed:
• at first use of the instrument,
• after relocating the instrument,
• after changing essential parts of the system, for example software updates or exchange
of cartridges,
• after instrument repair,
• on regular time intervals.
Verification Procedure
To perform verification tests:
1 Switch to the Verification context.
2 From the File menu select New.
3 A New Verification item appears in the Tree View Panel.
4 Under Cartridge Details, click Select and specify details on the cartridge that is currently
installed in the bioanalyzer.