Installation
40 Agilent 240/280 Series AA (including Zeeman)
10 Restart SpectrAA to save the configuration settings.
11 Go to the next section ‘Configuring the IEEE addresses for Agilent AA-DUO
operation’.
Configuring the IEEE addresses for Agilent AA-DUO operation
If you are running an Agilent AA-DUO system, you must reconfigure one of the
instruments so they have different IEEE addresses. Agilent 55B AA instruments
should have a default IEEE address of 1. Agilent 240FS AA, 240Z AA, 280FS AA
and 280Z AA instruments should have a default IEEE address of 9. Because 55B
AA’s must be Instrument #1 in the configuration dialog, Instrument #2 will always
have a default IEEE of 9. It is important not to change the IEEE address of a 55B
AA as it may cause conflicts with the inbuilt computer.
1 Close SpectrAA if it is open.
2 Navigate to C:\Program Files (x86)\Agilent\sp100\Run and then double-click
DUAL.EXE. Click
OK to start the program.
3 A window appears asking you to turn on the instrument you wish to
reconfigure. You must select ‘Instrument #2’ in the configuration. Switch on
Instrument #2 and then click
OK.
Do not turn on the Instrument #1.
A command window is displayed. The process may take about a minute.
During this time, you will hear the instrument resetting.
Once the process is complete, a message appears in the command window
saying ‘Addr changed from 9 to 8. Press any key to continue…’
If this message does not appear, continue the procedure below. If it does
appear, you have completed the setup.
4 In the search field type cmd.exe. Select cmd.exe from the list that appears.
5 Turn on Instrument #1 and Instrument #2, if not on.
6 Choose one of the following options:
For a Duo system with a 55B AA:
Type: CD C:\Program Files (x86)\Agilent\SP100\Run\ and then press
Enter.
Type: setaddr.exe q1:8 and then press
Enter.
For a Duo system without a 55B AA:
Type: CD C:\Program Files (x86)\Agilent\SP100\Run\ and then press

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 240 Series AA and is the answer not in the manual?

Agilent Technologies 240 Series AA Specifications

General IconGeneral
BrandAgilent Technologies
Model240 Series AA
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals