Maintenance and Troubleshooting
Agilent 240/280 Series AA (including Zeeman) 59
Disassembled spray chamber and burner where:
1. O-ring on burner assembly
2. O-ring inside the spray chamber
3. O-ring inside the bung
Spray chamber removed from gas inlet where:
1. O-rings (x2) that seal the gas inlets to the spray
chamber
Clean the lamp and sample compartment windows
Check the air compressor filter
Check the water levels in recirculating supplies (Furnace/Zeeman systems
only)
Yearly
Arrange for an Agilent service engineer to perform preventative maintenance.
Cleaning
Instrument surfaces
Clean contamination immediately using the approved laboratory procedures.
If there is contamination inside the instrument immediately turn off the instrument
and call your Agilent representative.
1 Switch off the instrument and allow it to cool.
2 Remove any hardware present in the sample compartment (refer to the
appropriate instructions in this User’s Guide and the SpectrAA Help if
necessary).
3 Using a cloth dampened with water, clean the sample compartment,
instrument cover, and chimney. On Flame systems, also clean the front panel
1
2
3
1