34410A/11A User’s Guide 125
Specifications 5
DC Voltage
Measurement Method: Continuously integrating multi–slope IV
10 VDC Linearity: 0.0002% of reading + 0.0001% of range
Input Resistance:
0.1 V, 1 V, 10 V Ranges Selectable 10 MW or >10 GW
(For these ranges, inputs beyond ±17 V are clamped through 100 kW
typical)
100 V, 1000 V Ranges 10 MW ±1%
Input Bias Current: < 30 pA at 25 °C
Input Terminals: Copper alloy
Input Protection: 1000 V
DC CMRR 140 dB for 1 k
W
unbalance in LO lead. ±500 VDC maximum
Resistance
Measurement Method: Selectable 4–wire or 2–wire ohms.
Max. Lead Resistance 10% of range per lead for 100
W
, 1 k
W
ranges.
(4–wire ohms) 1 k
W
per lead on all other ranges
Input Protection: 1000 V on all ranges
Offset Compensation: Selectable on the 100
W
, 1 k
W
, and 10 k
W
ranges
DC Current
Shunt Resistor: 0.1
W
for 1 A, 3 A.
2
W
for 10 mA, 100 mA.
200
W
for 100 µA, 1 mA.
Input Protection: Externally accessible 3 A, 250 V fuse
Continuity / Diode Test
Response Time: 300 samples / sec with audible tone
Continuity Threshold: Fixed at 10
W
Autozero OFF Operation (Typical)
Following instrument warm–up at a stable ambient temperature ±1 °C and <5 minutes.
Add 0.0002% of range + 2 µV for DCV or + 2 m
W
for resistance.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristics, and input
signal changes. Default delays are selected to give first reading right for most measurements.
Measurement Considerations
Agilent recommends the use of Teflon or other high–impedance, low–dielectric absorption wire
insulation for these measurements.

Table of Contents

Other manuals for Agilent Technologies 34411A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34411A and is the answer not in the manual?

Agilent Technologies 34411A Specifications

General IconGeneral
BrandAgilent Technologies
Model34411A
CategoryMultimeter
LanguageEnglish

Related product manuals