34410A/11A User’s Guide 71
Features and Functions 2
• The configured range mode (auto or manual) will be used for reading
hold.
• To enable reading hold, press to display TRIGGER.
Press as needed to select HOLD, and press .
Internal Triggering
The internal triggering mode is available from the remote interface only.
In the internal trigger mode, the trigger signal is always present. When
you place the multimeter in the wait–for–trigger state, the trigger is issued
immediately. This is the default trigger source for remote interface
operation.
• Remote Interface Operation: The following command selects the internal
trigger source:
TRIGger:SOURce IMMediate
The CONFigure and MEASure? commands automatically set the trigger
source to IMMediate.
Refer to the Agilent 34410A/34411A Programmer’s Reference for
complete description and syntax for these commands.
Software (Bus ) Triggering
The bus trigger mode is available from the remote interface only.
The bus trigger mode corresponds in function to pressing the key
from the front panel, except that it is initiated by sending a bus trigger
command, after selecting BUS as the trigger source.
• The following comand selects the bus trigger source:
TRIGger:SOURce BUS
Then, any of the following commands will initiate readings:
MEASure?
READ?
INITiate
Refer to the Agilent 34410A/34411A Programmer’s Reference for
complete description and syntax for these commands.