EasyManuals Logo

Agilent Technologies 34411A User Manual

Agilent Technologies 34411A
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
34410A/11A User’s Guide 129
Specifications 5
Frequency and Period
Measurement Type: Reciprocal–counting technique. AC–coupled input using the AC voltage
measurement function.
Input Impedance: 1 MW ±2%, in parallel with <150 pF
Input Protection: 750 V rms all ranges
Measurement Considerations
All frequency counters are susceptible to error when measuring low–voltage, low–frequency signals.
Shielding inputs from external noise pickup is critical for minimizing measurement errors.
Settling Considerations
Errors will occur when attempting to measure the frequency or period of an input following a dc offset
voltage change. The input blocking RC time constant must be allowed to fully settle ( up to 1 sec. )
before the most accurate measurements are possible.

Table of Contents

Other manuals for Agilent Technologies 34411A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34411A and is the answer not in the manual?

Agilent Technologies 34411A Specifications

General IconGeneral
BrandAgilent Technologies
Model34411A
CategoryMultimeter
LanguageEnglish

Related product manuals