80 34410A/11A User’s Guide
2 Features and Functions
• After the programmed count has been reached, the display will show
SAVING READINGS, then LOGGING COMPLETE. The readings are stored in
non–volatile memory.
To stop a data logging event in progress, you may press either ,
or . The primary display line will show DATA LOGGER? and
offers selections of CONTINUE or STOP. The data logger continues to take
readings while this is displayed.
• If you select STOP and press , the primary display will read DATA
TO NV?, prompted you to SAVE or DISCARD the readings already logged.
Select one of these options and press to terminate the data
logging function.
• If you select CONTINUE and press , the data logging function will
continue as programmed, until complete.
To read the data logged from the front panel, press to display
DATA LOGGER. Press to select VIEW–RDGS, and press . Press or
to scroll through the stored readings.
• Remote Interface Operation: The following command returns all readings
stored in non–volatile memory (NVMEM):
DATA:DATA? NVMEM
The following command deletes all readings stored in NVMEM.
DATA:DELete NVMEM
The following command returns the number of data points stored in
NVMEM
DATA:POINts? NVMEM
Refer to the Agilent 34410A/34411A Programmer’s Reference for
complete description and syntax for these commands.