Introduction
Agilent Microwave Plasma Atomic Emission Spectrometers User’s Guide 23
A chemically-inert container, not glass, to hold a minimum of 2 liters
(4 pints) of waste must be provided by the user. It should be located
underneath the sample tray (or on the right side of the MP-AES
instrument), where it is protected by the bench and in full view of the
operator.
Disposal of all chemical waste must comply with any rules and/or
regulations that may be imposed by the local authorities.
Personal computer requirements
The recommended and minimum PC specifications can be found in
the Agilent MP-AES Site Preparation Guide.
Locate the PC keyboard and mouse for ergonomically correct access.
Weights and dimensions
For the weights and dimensions of your MP-AES instrument please
see the Agilent MP-AES Site Preparation Guide.
WARNING
Heavy Weight
Danger to feet and hands. The Agilent MP-AES instrument weighs
approximately 73 kg (161 lb). To avoid injury to personnel or damage to
equipment, always use a fork lift or other suitable lifting device when moving
the MP-AES instrument.
Position the MP-AES instrument on the bench to allow easy access to
the power switch located on the right side of the instrument.