Operation
Agilent Microwave Plasma Atomic Emission Spectrometers User’s Guide 39
NOTE
You must install the correct peristaltic pump tubing for the wavelength
calibration:
Pump Drain
4100 MP-AES White/white Blue/blue
4200 and 4210 MP-AES Orange/green Blue/blue
2 Click the Instrument button.
3 Click Calibration.
4 Place the solution inlet tubing into the wavelength calibration
solution and allow the sample to reach the plasma.
5 Click Calibrate Instrument. The torch will be aligned and then a
wavelength calibration and a calibration check will be performed
automatically.
6 After a short while, an indication of the success or failure of the
calibration check will appear, as well as an indication of the
wavelength offsets.
7 If the calibration fails, check the sample introduction system. If
the system seems fine, prepare a new wavelength calibration
solution and try again, or consult the Troubleshooting section of
the MP Expert Help.
Creating/Opening a Worksheet
Creating a new worksheet
To create a new worksheet, click New from the Start page or the File
menu.
A list of recently used files will be presented when creating a new
worksheet from a template; otherwise you may Browse for more files.
The ‘New From Template’ dialog box will be displayed in this
instance.

Table of Contents

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals