- 20 -
Scientific Instruments Manufacturer GmbH
D-46149 Oberhausen +49-208-94 10 78-0 www.sim-gmbh.de
CONDITIONS for DIP/7000 Triple Q MS/MS
DIP Ready / Bereit
DIP method
check_EI_scan.dip (40°C for 1 min, 40-350°C at 1.0°/s, 350°C for 5
min)
MS/MS Ready
MS method DIP_EI_Scan.M
MS aux temp 280°C
Tune file atunes.ei.tune.xml
Ion Sorce EI
Aquisition mode SCAN
Scan Type MS1 Scan
Stop Mode By Stop Time
Stop time 10 min
Column flow 0.5 ml/min
Solvent delay 0.3 min
Time filter enabled yes
Time filter peak width
0.7 s
Use gain no
EMV Use tune setting (Delta EMV=0)
Mass range (low) 100.0
Mass range (high) 600.0
Scan time 100 ms
Threshold 0
Step size 0.1 amu
MS quad temp
(Q1 + Q2)
150°C
MS ion source temp :
230°C
Sample triglyceride standard
Push rod tip A: for liquid samples, B: for solid samples
Sample volume 1 µl
PROCEDURE
Set up the Methods
1. Load method DIP_EI_Q1.M for the 5975 MSD (MS ChemStation) or DIP_EI_Scan.M
for the 7000 Triple Q MS/MS.
2. Select the actual autotune using Instrument/Select MS Tune File (ATUNE.U or
atunes.ei.tune.xml).
3. Save the method to “…\1\METHODS\” as “DIP_EI_Q1.M” resp. “DIP_EI_Scan.M”
4. Print the method and verify that method and conditions are the same.
5. Load method “check_EI.dip” for the DIP-system (via DIP software, see DIP manual).

Other manuals for Agilent Technologies 5975 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 MSD and is the answer not in the manual?

Agilent Technologies 5975 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 MSD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals