The following steps have to be performed with the push rod tip for liquid samples (A) and the
tip for solid samples (B):
1. After loading the DIP method, click the Start button (green arrow) in the Instrument
Control view (of the MS software) and specify all sample information in the Sample
Information panel.
2. Click on “Start Run” and wait for the Ready-signal of the MS (“Acquisition: Waiting for
remote start …”).
3. If the Ready-signal appears in the status display (green field), you can inject the
sample according to the DIP manual (see 6.3) by pressing the “START” key to move
the push rod into the “Load” position.
4. Start the DIP using the controller or the “DIP”-software (see DIP manual, 6.2): Click
on the “START” key of the controller or in the “DIP” programm to start the
measurement.
RESULTS
EI PERFORMANCE
Item
t
max
(TIC)
[min]
Typical
value
[min]
Main signals
[m/z]
Typical values
[m/z]
Abundance (TIC, t
max
)
[counts]
Typical value
[counts]
Tricaprin
1.5-4.0
155/229/383
> 1 000 000
Tripalmitin
4.5-7
239/313/551
> 600 000
Example for DIP/5975MSD System:

Other manuals for Agilent Technologies 5975 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 MSD and is the answer not in the manual?

Agilent Technologies 5975 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 MSD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals