EasyManuals Logo

Agilent Technologies 6460 Guide

Agilent Technologies 6460
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
Exercise 3 – Create a Triggered Dynamic MRM acquisition method
Task 2. Add/Modify compounds in an existing database
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide 45
Task 2. Add/Modify compounds in an existing database
You can also manually add compounds to a database and modify the
compounds in the database. In the next task, you create a Triggered Dynamic
MRM method from the compounds in the database.
Steps Detailed Instructions Comments
1 Review the
iiiSulfamix_dMRM2.m, where iii
are your initials.
a Click File > Open > Method.
b Select the iiiSulfamix_dMRM2.m
method.
c Click OK.
d Review the parameters.
A Triggered Dynamic MRM method
is a type of Dynamic MRM method.
The Scan Type for both methods is
Dynamic MRM.
2 Start the MassHunter Optimizer
program.
Double-click the Optimizer icon. . If you are optimizing peptides, use
the Optimizer for Peptides program.
3 Set parameters on the Optimizer
Setup tab.
a Click the Optimizer Setup tab.
b Click the Injection (with or without
column) button.
c Set the CE range from
4 to 48.
d Set the Cell Accelerator Voltage to
7.
e Right-click the table and click Add
Method.
f Select the iiiSulfamix_dMRM2.m
method.
To create low mass product ions
from a precursor ion near 300 m/z,
you need fairly high collision
energies.
4 Set parameters on the Precursor
Ion Selection tab.
a Click the Precursor Ion Selection tab.
b Verify that +H is marked for the
Positive ions (with priorities) list.
5 Set parameters on the Product Ion
Selection tab.
a Click the Product Ion Selection tab.
b Click the Mass (m/z) button under
Low mass cut-off.
c Enter
60 for the low mass cut-off.
On the Product Ion Selection, you
can automatically add up to 4
product ions per compound (for
instance, 2 primaries and 2
secondaries). You want 8 to 10
peaks in the composite spectrum to
prove that this is indicative of the
compound, so you need to add at
least some of the product ions
manually.

Other manuals for Agilent Technologies 6460

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6460 and is the answer not in the manual?

Agilent Technologies 6460 Specifications

General IconGeneral
BrandAgilent Technologies
Model6460
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals