Operation Guide
Syringes
17
7. Move the plunger carrier loop down and tighten the plunger screw.
8. Move the plunger carrier loop up and down. If the syringe plunger does not
move along with the carrier, repeat the previous steps. Be sure the plunger
screw is tight.
9. Check that the needle is aligned with the needle guide in the foot by moving
the slide up and down. The needle should slide smoothly in the needle guide.
See Figure 11.
Figure 11 Syringe carriage and needle support with needle installed
Plunger screw
Plunger carrier loop
Flange guide
Syringe latch
Syringe clip
Needle support foot
Slide
Syringe carriage
Syringe
Needle
Needle support