56 Agilent Gas Sample Selector for 8860 and 8890 User Manual
4 Operation
Building a method
Term definitions
Load Time: The time for the sample to fill the sample loop.
Equilibration Time: After filling the sample loop, the GC will wait for this time to
allow the sample to equilibrate before starting the injection.
Purge Time… at: After the injection starts, and after the delay for the time entered
in the at field, the Selector will purge the GC sample loop with the purge gas for the
specified number of minutes.