Agilent B1500 VXIplug&play Driver Users Guide, Edition 3 2-13
Driver Functions
Table 2-7 MFCMU measurement range
range
Measurement range
a
a. Available measurement ranges depend on the measurement frequency.
200 kHz 2 MHz 5 MHz
range = 0 auto ranging
0 < range < 100 50 Ω 50 Ω 50 Ω
100 range < 300 100 Ω 100 Ω 100 Ω
300 range < 1000 300 Ω 300 Ω 300 Ω
1000 range < 3000 1 kΩ 1 kΩ 1 kΩ
3000 range < 10000 3 kΩ 3 kΩ 3 kΩ
10000 range < 30000 10 kΩ 10 kΩ
30000 range < 100000 30 kΩ 30 kΩ
100000 range < 300000 100 kΩ
300000 range 300 kΩ