2-26 Agilent B1500 VXIplug&play Driver User’s Guide, Edition 3
Driver Functions
agb1500_execLoadCorr
channel Slot number of the slot that installs the MFCMU to be used. See
Table 2-2. channel=-1 detects the slot automatically.
result Execution result of this function. Returned value.
0: Passed. No failure detected.
1: Failed.
2: Aborted.
agb1500_execLoadCorr
This function performs the CMU load correction for the specified measurement
frequency, and sets the correction data to the B1500. This function also returns the
execution results. This function initializes the MFCMU.
To execute this function, connect the load standard that has the reference value or
calibration value. And enter the values to the primary and secondary parameters to
set the reference value to the B1500.
The agb1500_forceCmuAcLevel function must be executed before this function.
Syntax ViStatus _VI_FUNC agb1500_execLoadCorr(ViSession vi, ViInt32 channel,
ViReal64 freq, ViInt32 mode, ViReal64 primary, ViReal64 secondary,
ViPInt16 result);
Parameters vi Instrument handle returned from agb1500_init( ).
channel Slot number of the slot that installs the MFCMU to be used. See
Table 2-2. channel=-1 detects the slot automatically.
freq Frequency (in Hz). 1000 (1 kHz, initial setting) to 5000000
(5 MHz). Setting resolution: 1 mHz (1 kHz to), 10 mHz
(10 kHz to), 0.1 Hz (100 kHz to), or 1 Hz (1 MHz to 5 MHz).
mode MFCMU measurement parameters. This value must be 400 at
this time.
primary Reference value of the standard. The value must be the value
for the primary parameter (ex: R in the R-X measurement
mode). Numeric expression.
secondary Reference value of the standard. The value must be the value
for the secondary parameter (ex: X in the R-X measurement
mode). Numeric expression.
result Execution result of this function. Returned value.

Table of Contents

Other manuals for Agilent Technologies B1500A

Questions and Answers: