Agilent B1500 VXIplug&play Driver Users Guide, Edition 3 2-41
Driver Functions
agb1500_sampleIv
agb1500_sampleIv
This function executes a sampling measurement by the specified channels, and
returns the number of measurement points, measurement data index, measurement
data, measurement status, and time stamp data.
Before this function, execute the agb1500_setSample function to set the sampling
timing and execute the agb1500_setSampleMode function to set the sampling mode.
Also, execute the agb1500_addSampleSyncIv functions to set the synchronous DC
sources used with the sampling measurement channels.
Syntax ViStatus _VI_FUNC agb1500_sampleIv(ViSession vi, ViInt32 channel[ ],
ViInt32 mode[ ], ViReal64 range[ ], ViPInt32 point, ViInt32 index[ ],
ViReal64 value[ ], ViInt32 status[ ], ViReal64 time[ ]);
Parameters vi Instrument handle returned from agb1500_init( ).
channel[ ] Slot number of the slot that installs the SMU to be used. See
Table 2-2.
Enter 0 to the last element of channel[ ]. For example, if you use
two channels, set the array size to 3, specify the channels to the
first and second elements, and enter 0 to the third element.
mode[ ] Measurement mode. 1 (current) or 2 (voltage).
range[ ] Measurement ranging mode. 0 (auto), positive value (limited
auto), or negative value (fixed range). For the available values,
see Table 2-3.
For the linear sampling with interval < 2 msec, only the fixed
range is available. Even if range=0 or positive value is set, the
measurement channels will automatically select and use the
range that covers the compliance value set to each channel.
point Number of measurement points. Returned data.
index[ ] Measurement data index. Returned data.
value[ ] Measurement data. Returned data.
status[ ] Measurement status. Returned data. For the status value, see
“Status Code” on page 2-14. To disable the status data output,
set 0 (NULL pointer) instead of array.
time[ ] Time stamp data (measurement start time). Returned data. To
disable the time stamp data output, set 0 (NULL pointer)
instead of array.

Table of Contents

Other manuals for Agilent Technologies B1500A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies B1500A and is the answer not in the manual?

Agilent Technologies B1500A Specifications

General IconGeneral
BrandAgilent Technologies
ModelB1500A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals