2-60 Agilent B1500 VXIplug&play Driver User’s Guide, Edition 3
Driver Functions
agb1500_setSerRes
mode Sampling mode, linear or logarithm.
0: linear sampling
1: logarithmic sampling, 10 data/decade.
2: logarithmic sampling, 25 data/decade.
3: logarithmic sampling, 50 data/decade.
4: logarithmic sampling, 100 data/decade.
5: logarithmic sampling, 250 data/decade.
6: logarithmic sampling, 500 data/decade.
agb1500_setSerRes
This function sets the series resistor of the specified channel.
Syntax ViStatus _VI_FUNC agb1500_setSerRes(ViSession vi, ViInt32 channel,
ViInt32 state);
Parameters vi Instrument handle returned from agb1500_init( ).
channel Slot number of the slot that installs the SMU to be used. See
Table 2-2. Set 0 to select all SMUs.
state 0 (disconnects resistor) or 1 (connects resistor).
agb1500_setShortCorrMode
This function sets the CMU short correction function ON or OFF.
The agb1500_execShortCorr function must be executed before this function.
Syntax ViStatus _VI_FUNC agb1500_setShortCorrMode(ViSession vi, ViInt32 channel,
ViInt32 mode);
Parameters vi Instrument handle returned from agb1500_init( ).
channel Slot number of the slot that installs the MFCMU to be used. See
Table 2-2. channel=-1 detects the slot automatically.
mode 0: CMU short correction function ON.
1: CMU short correction function OFF.

Table of Contents

Other manuals for Agilent Technologies B1500A

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies B1500A and is the answer not in the manual?

Agilent Technologies B1500A Specifications

General IconGeneral
TypeSemiconductor Device Analyzer
ModelB1500A
Maximum Voltage200 V
Number of SMUsUp to 10
Frequency RangeDC to 1 MHz
ConnectivityUSB, LAN
ChannelsUp to 10
InterfaceGUI

Summary

Installation

System Requirements

Details the necessary hardware and software for the driver.

Installing VXIplug&play Driver

Provides a step-by-step guide for installing the driver.

Before Programming

Covers pre-programming setup and configuration steps.

Driver Functions

Function List

Lists all available driver functions with brief descriptions.

Parameters

Details the parameters used by various driver functions.

Status Code

Describes the status codes returned by measurement functions.

Function Reference

Provides detailed descriptions and syntax for each driver function.

Visual Basic Programming Examples

Programming Basics

Basic information and steps for programming with the driver.

High Speed Spot Measurement Example

Example demonstrating high-speed spot measurement using Visual Basic.

Staircase Sweep Measurement Example

Example for staircase sweep measurements.

Breakdown Voltage Measurement Example

Example for measuring breakdown voltage.

Leakage Current Measurement Example

Example for measuring leakage current.

C++ Programming Examples

Programming Basics

Basic information and steps for C++ programming.

High Speed Spot Measurement Example

Example demonstrating high-speed spot measurement in C++.

Staircase Sweep Measurement Example

Example for staircase sweep measurements in C++.

Breakdown Voltage Measurement Example

Example for measuring breakdown voltage in C++.

Leakage Current Measurement Example

Example for measuring leakage current in C++.

Sampling Measurement Example

Example for sampling measurement in C++.

CV Sweep Measurement Example

Example for CV sweep measurement in C++.

Related product manuals