2-34 Agilent B1500 VXIplug&play Driver User’s Guide, Edition 3
Driver Functions
agb1500_msweepIv
agb1500_msweepIv
This function performs sweep measurement, and returns the number of
measurement steps, sweep source data, measurement data, measurement status, and
time stamp data.
Before this function, execute the agb1500_setIv function to set the primary sweep
source and execute the agb1500_setNthSweep function to set an synchronous sweep
source. Up to nine synchronous sweep sources can be set by using the
agb1500_setNthSweep function for each channel.
Syntax ViStatus _VI_FUNC agb1500_msweepIv(ViSession vi, ViInt32 channel,
ViInt32 mode, ViReal64 range, ViPInt32 point, ViReal64 source[ ],
ViReal64 value[ ], ViInt32 status[ ], ViPReal64 time[ ] );
Parameters vi Instrument handle returned from agb1500_init( ).
channel Slot number of the slot that installs the SMU to be used. See
Table 2-2.
mode Measurement mode. 1 (current) or 2 (voltage).
range Measurement ranging mode. 0 (auto), positive value (limited
auto), or negative value (fixed range). For the available values,
see Table 2-3.
point Number of measurement steps. Returned data.
source[ ] Sweep source setup data. Returned data. To disable the source
setup data output, set 0 (NULL pointer) instead of array.
value[ ] Measurement data. Returned data.
status[ ] Measurement status. Returned data. See “Status Code” on page
2-14. To disable the status data output, set 0 (NULL pointer)
instead of array.
time[ ] Time stamp data (measurement start time). Returned data. To
disable the time stamp data output, set 0 (NULL pointer)
instead of array.
Example
ViSession vi;
ViStatus ret;
ViInt32 emitter = 1; /* SMU1 */
ViInt32 base = 2; /* SMU2 */
ViInt32 collector = 4; /* SMU4 */
ViReal64 vb1 = 0.25;
ViReal64 vb2 = 0.75;
ViReal64 vc = 3;

Table of Contents

Other manuals for Agilent Technologies B1500A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies B1500A and is the answer not in the manual?

Agilent Technologies B1500A Specifications

General IconGeneral
BrandAgilent Technologies
ModelB1500A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals