28 Chapter 2
Troubleshooting
Booting Process Troubleshooting
Figure 2-11 Firmware revision and option information
If the display whited out, entirely blued or appeared with a dialog box, a mass storage
problem is suspected. Try to perform the mass storage recovery procedure.
NOTE If a message of "Will Shut Down in Five Seconds" is displayed in place of "Initializing.."
and the shutdown occurs, the DSP board fails in starting up. The following message may
be displayed before the shutdown occurs:
"Fatal Error: Failed to Initialize DSP Driver":
or "Fatal Error: Failed to Initialize DSP":
This message indicates that the DSP board doesn't work or is not properly connected to the
DSP board.
"Fatal Error: Failed to Update DSP Code":
If this happened, the DSP board failed in writing DSP program into flash ROM when the
firmware was installed first or updated to the newest version. A problem in the DSP or A50
CPU Module is suspected.
Step 5. Measurement view
The measurement view as shown in Figure 2-12 is displayed after the system initialization
is completed without problem.

Table of Contents

Other manuals for Agilent Technologies E5061B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5061B and is the answer not in the manual?

Agilent Technologies E5061B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5061B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals