EasyManua.ls Logo

Agilent Technologies E5061B - Page 46

Agilent Technologies E5061B
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 45
Troubleshooting
Performance Test failure Troubleshooting
2. Troubleshooting
assembly and others are tight or loose. Also check for possible disconnection (impairment)
of the cables and connectors.
Table 2-8 Performance Test failure Troubleshooting information (Option 3L5)
Test Num.
First failed test Probable faulty board assembly or
parts
A1/A11
Source
Module
A2
Receiver
Module
A9 LF
Source
Bias
Module
1 Frequency accuracy test ### # #
2 RF output level accuracy and
flatness test
### # #
3 RF output level linearity test ### # #
4 Trace noise test ## ### #
5 Crosstalk & system Dynamic
Range test
## ### #
6 Dynamic accuracy test # ### #
7 Uncorrected system performance
test
## ### #
8 LF output level accuracy& flatness
test (Opt 3L5)
## ###
9 LF output level linearity test (Opt
3L5)
## ###
10 DC bias accuracy test (Opt 3L5) # ###
11 Crosstalk & noise level test (Opt
3L5)
# ### #
12 Input impedance test (Opt 3L5) ###
13 Absolute amplitude accuracy test
(Opt 3L5)
# ### #
14 Magnitude ratio/phase dynamic
accuracy test (Opt 3L5)
# ### #
15 Magnitude ratio/phase frequency
response test (Opt 3L5)
# ### #
16 Trace noise test (Opt 3L5) ## ### #
17 DC measurement accuracy test
(Opt 3L5)
### ###

Table of Contents

Other manuals for Agilent Technologies E5061B