2-76
Assembly Replacement
A28 YIG Oscillator
A28 YIG Oscillator
Tools Required
T-10 driver
5/16” open ended wrench
Removal Procedure
Refer to Figure 2-38 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the A31 Motherboard facing you and the RF deck on top.
4. Using the 5/16’ open ended wrench, disconnect the W38 RF cable from the A28 YIG Oscillator.
5. Disconnect the W12 ribbon cable from the A28 YIG Oscillator’s driver board.
6. Using the T-10 driver, remove the two screws (1) that attach the A28 YIG Oscillator to the side
panel.
7. Using the 5/16” open ended wrench, disconnect the W27 RF cable from the A30 Modulation
Filter with High Power.
8. Remove the A28 YIG Oscillator.
Replacement Procedure
Reverse the order of the removal procedure.
Torque all T-10 screws to 9 in-lbs.
Torque all RF connectors to 8 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8267C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267C and is the answer not in the manual?

Agilent Technologies E8267C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267C
CategoryPortable Generator
LanguageEnglish

Related product manuals